Home>Equipment>Rental>Kaye Instruments> Kaye Validator 2000

Kaye Validator 2000

Kaye Validator 2000

$0.00

A thermal validation solution for inspection of pharmaceutical, biotechnology and medical device manufacturing. Recognized and widely used Validation System for the Pharmaceutical & Biotech Industry for wired application

Thank you for your request, we will get back to you shortly

Subscribe to receive our latest updates!

Request to rent Download Manual

We will get back to you shortly

Please enter your email address

Description

A thermal validation solution for inspection of pharmaceutical, biotechnology and medical device manufacturing. Recognized and widely used Validation System for the Pharmaceutical & Biotech Industry for wired application
Features
  • Intuitive study setup and extensive process calculations simplify valdiation process
  • Improved data analysis, including ability to graph and report unlimited cycles and up to 25 groups
  • Centralized network management of all passwords and audit trails
  • Ability to merge data files from multiple Kaye Validator and ValProbe studies
  • Sensor input modules minimize sensor handling and calibration, including 4/20 mA inputs
  • Fully automated sensor calibration
  • Operates standalone or with PC
  • Enables compliance with FDA 21CFR Part 11 guidelines
  • Total system accuracy, following guidelines for uncertainty measurement of 0.17°C (k=2)
Kaye Validator
resolution and short term stability
0.017°C
k=1
IRTD Temperature Standard 0.01°C
k=1
Temperature Reference 0.051
K=1
Total System Uncertainty 0.078°C
k=1
Kaye Validator Specifications
Analog Input up to 36
Thermocouples Type T, J, K: 0.1°C or F resolution; T limited range 0.01°C resolution
Scanning Speed 8 or 12 inputs/sec at 50/60 Hz
Internal Memory 2 Mb for data collection
Internal Battery NiCad; 30 minutes of battery backup
Input Indepedance 10KΩ. Source greater than 10KΩ produces open circuit indication
Common Mode
Rejection
160 db (8 inputs/sec) @ line frequency
145 db (12 inputs/sec) @ line frequency
140 db @ DC
Max. Common
Mode Voltage
100V pk ch-to-ch
350V pk ch-to ch to frame ground
Normal Mode
Rejection
82 db @ 60 Hz (8 inputs/sec)
69 db @ 60 Hz (12 inputs/sec)
Voltage Input 0 to 10 VDC
Resolution 1:72,000
Voltage Input
Accuracy
30 days: ±(0.003% of reading + 2 counts + 4 microvolts)
1 year: ±(0.006% of reading + 2 counts + 4 microvolts)
Sensivity 0.5 microvolts/count on most sensitive range
Voltage Temp.
Coef.
±(0.1 microvolts + 0.001% reading)/°C
Compensator
Temp Coef.
±0.01°C per °C
Input Terminal Temperature
Non-unformity
0.1°C
Input Ranges -6 to 30mV, -60 to 300mV, -2 to 10V
Environmental Temperature: 0 to 50°C (32 to 122°F)
Relative humidity: 95% non-condensing
Power 90 to 250 VAC, 50/60 Hz
Fuse Rating 2A Slo Blo
Size 191 H x 343 W x 343 mm D (404 mm with SIM)
7.5 in H x 13.5 in W x 13.5 in D (15.9 in with SIM)
Weight 9 kg (20 lbs)
  • Steam Sterilizers (Autoclaves)
  • Dry Heat Sterilizers
  • Steam in Place (SIP)
  • Water Cascade/Fall Sterilizers
  • Incubators
  • Stability Chambers
  • Freezers
  • Freeze Dryer/Lyophilisation
  • Vessel